4 INSPECTION AND FUNCTIONAL TESTING
Safety measures as described in Section 1 must be observed when inspecting or testing a chair. Only authorized
caregivers or maintenance staff should attempt to inspect or test a chair.
4.1
Inspection
We recommend regular visual inspection for signs of wear, damage, loose or missing fittings, and other safety
concerns. If a breakage, defect, or operational problem is detected, the chair must be repaired inspected and tested
for function before it is returned to service.
We recommend that the chair be inspected as often as each use if there is any reason to be concerned about the
possibility of increased wear or loose or missing fittings. At a minimum, in regular use the chair should be
inspected on a bimonthly basis. The visual inspection procedure should include at least the observation of all the
fittings (fasteners):
The visual inspection procedure should include at least the observation of the following parts:
1)
The cylinder that positions the seat tilt
2)
The vinyl straps used in the seat, back and wings
3)
The cushions
4)
The black handle and cable
5)
The armrest height adjustment pins
The visual inspection procedure should include the observation of any installed accessories. The visual inspection
should be performed by the facility, or if in a private residence, by the individual responsible for the chair.
4.2
Functional Testing
We recommend that the chair should be tested for operation of the chair's functions without a resident in the
chair. The testing may be as often as each use if there is any reason to be concerned about the possibility of
increased wear or damage to the chair's components. At a minimum, in regular use the chair should be tested for
functions as described in Section 3.
If the caregiver or maintenance department performing the functional testing believes that any function is not
operating correctly, the chair should be taken out of service until a satisfactory functional test can be completed.
The caregiver or maintenance department performing the testing should be aware that the seat tilt will be
more difficult without a resident in the chair. When performing repairs or maintenance, do not use lubricants that
contain solvents. Solvents will damage many of the moving components in the chair. If necessary, a white, food
grade grease (lubricant) may be used on the sliding components in the chair. Do not use spray lubricants on any part
of the chair.
15