ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer's declaration - electromagnetic immunity
The TS42S IR Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the TS42S IR Thermometer should assure that it is used in
such an environment.
Emissions Test IEC 60601 Compliance Electromagnetic
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
24 • English
Table 2 EMISSIONS
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Environment Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.