Error message
13.2.4 Status signals
Battery charger is connected
LED on
LED on bat
power
tery charger
supply
Battery charger disconnected
Beep
Vibration
signal
signal
1x short
1x short
3x short
Battery charge level
Battery char
ger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
Battery is fully charged (or connection with product is interrupted).
Take note of the beep signals to differentiate between these states.
When the battery charger is connected or disconnected, a self-test is conduc
ted and confirmed by a single beep and vibration signal.
The battery is fully charged if this beep signal is heard.
If no beep signal is heard, the connection to the product is interrupted.
13.3 Technical data
Environmental conditions
Storage and transport in original packaging
(≤3 months)
68 | Ottobock
Cause
A current connection to
the prosthesis was inter
rupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery
charger. If the beep signal sounds again, visit your prosthetist soon. If
necessary, the prosthetist will forward the product to an authorised
Ottobock Service Centre.
The product can be used without restrictions. However, vibration signals
may not be generated.
Correction
Check the following points:
•
Distance from the prosthesis to the
device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched
on? (see Page 60)
•
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
-20°C/-4°F to +40°C/+104°F