2.4 Connections

1
2
3
4
5
6
7
8
9
10
11

2.5 Calibration cavities

You can use the 0.2 ml, 0.5 ml, 2.0ml and 5 ml for validity check of the probe calibration.
To perform a calibration check, select a protocol the measures a tympanogram.
Do not use an ear tip! Place the probe tip completely into the cavity. Perform the measurement. Check the
volume that was measured.
The allowed tolerance in the volume measurement is ±0.1 ml for cavities up to 2 ml and ±5% for larger
cavities. These tolerances are applicable for all probe tone frequencies.
We strongly recommend calibrating probe and contra phone at least once a year.
Probe
Dedicated probe connection
LAN
LAN (Not used)
USB B
For PC connection
USB A
For printer, mouse, keyboard, Memory stick
HDMI
For external monitor or projector
In 24 V
Use only specified power supply unit UE60 type
Trigger in / out
Cochlear implant trigger option
Pat. Resp.
Patient response button
Right
Audiometry Right output
Left
Audiometry Left output
Contra
Contra transducer output