To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use only
the following accessories:
ITEM
Power supply unit UE60
Clinical Probe
Diagnostic prob
DD45C Contra Headset DD45C
contra headset P3045
IP30 Insert Phone 10ohm single
contraEARTone 3A with Minijack
DD45 Audiometric Headset P3045
IP30 Insert Phone 10ohm set
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types and cable
lengths are as specified below:
Description
Mains Cable
USB Cable
Clinical Probe
Diagnostic probe
DD45C contra headset P3045
IP30 Insert Phone 10ohm single
contra
DD45 Audiometric Headset P3045
IP30 Insert Phone 10ohm set
MANUFACTURER
Interacoustics
Interacoustics
Interacoustics
Interacoustics
Interacoustics
Interacoustics
Interacoustics
Length
2.0m
2.0m
2.0m
2.0m
0.4m
0.5m
0.5m
0.5m
MODEL
UE60-240250SPAx
Clinical probe system 1077
Diagnostic probe system 1077
DD45C
Ear3AIP30C
DD45
IP30
Screened?
Unscreened
Screened
Unscreened
Unscreened
Screened
Screened
Screened
Screened
Page 55