Electromagnetic Compatibility (Emc) - Interacoustics AT235 Manual De Instrucciones

Ocultar thumbs Ver también para AT235:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

AT235 Instructions for Use - EN

5.4 Electromagnetic Compatibility (EMC)

Portable and mobile RF communications equipment can affect the AT235. Install and operate the AT235 according to the EMC
information presented in this chapter.
The AT235 has been tested for EMC emissions and immunity as a standalone AT235. Do not use the AT235 adjacent to or stacked with
other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by Interacoustics
as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
Guidance and manufacturer's declaration - electromagnetic emissions
The AT235 is intended for use in the electromagnetic environment specified below. The customer or the user of the AT235
should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the AT235.
The AT235 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the AT235 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the AT235 as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The AT235 is intended for use in the electromagnetic environment specified below. The customer or the user of the AT235 should assure
that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d = 1.17 √ ��
0.12
0.37
1.17
3.70
11.70
IEC 60601 Test
Level
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5% UT
(>95% dip in UT) for 0.5 cycle
40% UT
(60% dip in UT) for 5 cycles
70% UT
Electromagnetic environment - guidance
The AT235 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The AT235 is suitable for use in all commercial, industrial, business,
and residential environments.
80 MHz to 800 MHz
d = 1.17 √ ��
0.12
0.37
1.17
3.70
11.70
Compliance
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT) for
5 cycles
70% UT (30% dip in UT) for
Page 53
800 MHz to 2.5 GHz
d = 2.23 √ ��
0.23
0.74
2.33
7.37
23.30
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be greater than 30%.
Mains power quality should be that of a
typical commercial or residential
environment.
Mains power quality should be that of a
typical commercial or residential
environment.
Mains power quality should be that of a
typical commercial or residential
environment. If the user of the AT235
requires continued operation during
power mains interruptions, it is
recommended that the AT235 be

Publicidad

Tabla de contenido
loading

Tabla de contenido