Important Control Solution Testing Notes

Use only MediSense Glucose and Ketone control solutions with the meter
Replace the cap securely on the bottle immediately after use
IMPORTANT: Control solution results should fall within the control solution
range printed on the test strip instructions for use. Check that the lot number
printed on the test strip packet and instructions for use match.
Repeat the test if control solution results are outside this range
Stop using the meter if control solution results are continually outside the range printed on the
test strip instructions for use. Contact Customer Service.
IMPORTANT: Control solution results do not reflect the blood glucose or
blood ketone level.
Do not use control solution past the expiration date.
Do not add water or other liquid to control solution.
Do not swallow or inject the control solution.
Do not use the control solution as eye drops.
26