• Incoming water temperature to the Cavitron System should not
exceed 25˚C (77˚F). If needed a device should be installed to
maintain a temperature within this specification, or a Cavitron
DualSelect Dispensing System attached to allow this system to be
operated as a closed water system.

7.2 Electrical Requirements

• Incoming power to the system must be 100 volts AC to 240 volts
AC, single phase 50/60 Hz capable of supplying 1.0 amps.
• The system power should be supplied through the AC power cord
provided with your system.
• WARNING: To avoid risk of electric shock, this equipment must
only be connected to a supply mains with protective earth.

7.3 Unpacking the System

Carefully unpack your Cavitron Touch Ultrasonic Scaling System and
verify that all components and accessories are included:
1. Cavitron Touch System with Detachable Featherweight
Handpiece Cable
2. Water Line Assembly (Blue) with Filter and Quick Disconnect
3. Additional Water Line Filter
4. Detachable AC Power Cord
5. Cavitron® Tap-On™ Wireless Foot Pedal
6. Auxiliary Cable for Rechargable Foot Pedal
7. Steri-Mate® 360 Detachable Sterilizable Handpiece
8. Literature Packet

7.4 System Installation

• The Cavitron Touch System is designed to rest on a level surface.
Be sure unit is stable and resting on four feet.
• The Cavitron Touch System should not be positioned such that
access to the power input and AC power cord are limited.
• Placing unit in direct sunlight may discolor plastic housing.
• The system has been equipped with a Tap-On Wireless Foot Pedal
which was factory synchronized to operate with the system's base
unit. If your office has more than one Cavitron system with Tap-On
Technology, it is recommended that you mark the Tap-On Foot
Pedal and base unit for easy reference as to which Tap-On Foot
Pedal operates with which base unit. Should resynchronization be
necessary, follow the instructions in section 7.7.

7.5 Power Cord Connection

• Verify the Main Power ON/OFF switch, located at the rear of the
System, is set to the OFF (O) position before proceeding.
• Insert the AC power cord into the power input on the back of the
System.
• Insert the pronged plug into an AC wall outlet.

7.6 Water Supply Line Connection

• Grasp the Water Supply Line (blue hose) by the end opposite the
quick-disconnect and insert it into the water inlet connector until
fully seated.
• Connect the quick-disconnect to the dental office water supply or
a Cavitron DualSelect Dispensing System.
• Inspect all connections to make certain there are no leaks.
• To remove the water line from the Cavitron System, turn off the
dental office water supply. Disconnect the water supply line from
the dental office water supply. If a quick-disconnect connector
is attached to the end of the hose, relieve the water pressure by
pressing the tip of the connector into an appropriate container
and allow water to drain. To remove the hose from the system,
push on the outer ring of the system's water inlet and gently pull
out the water line.
7.7 Tap-On™ Foot Pedal Synchronization
The Tap-On Wireless Rechargeable Foot Pedal supplied with your
system has been factory synchronized with the base unit. Should
a replacement Foot Pedal be necessary, synchronization will be
required prior to system operation. Perform the following steps to
synchronize the Foot Pedal with the base unit.
Main Power Switch
Power Input
5
1. Turn the Main Power switch located at the rear of the system
to the ON (I) position. The Main screen will appear. This is also
referred to as the Scale screen.
2. Press and hold the Settings icon until the Settings screen
appears.
3. Maintain a distance of no more than 10 feet (3 meters)
between the base unit and Tap-On Foot Pedal during the
synchronization process.
4. Remove any inserts from the handpiece.
5. Ensure the foot pedal power is ON.
6. Press the Sync icon on the Settings screen.
7. Press the red sync button on the bottom of the foot pedal -
hold for at least three seconds.
8. Synchronizing is complete when the Sync icon has stopped
rotating and sound indication occurs.
9. To verify proper communication, press the Scale icon to return
to the Main screen (scale screen). Depress the foot pedal to
the second position and ensure that Boost is activated (seen
in Power Level icon).
Press ring to release
water supply tube.