Operation Status - Siemens THE PROBE Instrucciones De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Calibration
The calibration of the mA output may be done such
that its span will be either proportional or inversely
proportional to the material level.
The 4 and 20 mA levels may be calibrated in
Note:
any order.
Calibration: Reference Method
• Adjust the material level (or target) to the desired distance from the sensor face.
• Press the "4" or "20" key (as appropriate) to view the stored distance associated with that mA
output value.
• Press the key a second time to set the new distance reference.
• After viewing or calibrating, Probe operation automatically reverts to the Run mode (6 sec). The
calibration value is referenced from the face of The Probe sensor, in the units displayed.
4 mA calibration
Press "4"
Press "4" again
20 mA calibration
Press "20"
Press "20" again
Calibration bypasses the measurement response rate.
Note:

Operation Status

The graphic portion of the display gives the user a visual
indication of The Probe's operating status. Viewing the
graphic can assist the user in properly locating and
installing The Probe to achieve optimum performance.
The logo will change from full to partial to indicate operation status. After the `Waiting' period, the `?'
icon will appear for an `LOE / FAULT' indication. When a valid echo is again received, a `Good'
indication will resume. Refer to Troubleshooting on page 9 .
Page 4
proportional span
high level = 20 mA
low level = 4 mA
4 mA calibration
new 4 mA calibration
20 mA calibration
new 20 mA calibration
The Probe – INSTRUCTION MANUAL
inversely proportional span
high level = 4 mA
low level = 20 mA
calibration invalid if
retry
calibration invalid if
retry
GOOD
WAITING
LOE / FAULT
7ML19985GD62

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

The probe

Tabla de contenido