End Of Operation; Troubleshooting - Dräger 8413417 Instrucciones De Uso

Ocultar thumbs Ver también para 8413417:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
English / US English

End of operation

1 Switch off the connected ventilator.
Switching off the compressor
2 Press the On/Off switch on the rear (position
(off)).
The green LED
goes out.
Venting the compressor
The compressor is not vented automatically.

Troubleshooting

Fault
Compressor vibrates strongly
Compressor runs but builds up
too little or no pressure
Orange LED Temp
and a continuous buzzing
sounds
Compressor does not run
Orange LED Temp
and a continuous buzzing
sounds.
Compressor refuses to start
28
Cause
Rubber feet on compressor unit
defective
Intake filter soiled
Leakages in the medical air
compressor
lights up
High ambient temperature
Filter mat soiled
Fan defective
Device defective
lights up
Mains voltage too low or not
available
Fuses defective
Device defective
3 For devices with standby mode, remove the
connector from the wall terminal unit of the
central gas supply system (Air).
CAUTION
Only unscrew the compressed gas hose after
removing the connector from the wall terminal unit
of the central gas supply system. The
compressed gas hose is otherwise still
pressurized, which can injure the user.
4 Carefully remove the Air compressed gas hose
from the ventilator to vent the device.
The remaining compressed air escapes audibly.
Remedy
Call DrägerService
Perform maintenance. Call
DrägerService
Call DrägerService
Place the compressor in a cool
location and ensure free air
circulation
Knock dirt out of filter mat or
replace with a new filter mat
Call DrägerService
Call DrägerService
Check mains voltage
Call DrägerService
Call DrägerService
Instructions for Use Air Compressor