6. Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display reacts.
You can set damping to suit requirements.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
 fast determination of weight
1 = average damping
 normal determination of weight
2 = strong damping
 slow determination of weight.
– Press an arrow key if you want to
change the set value.
– Press the FUNC key again to store the
displayed value. The scale is then back
in normal weighing mode.
The settings are retained even once the scale is switched off.
7. Cleaning
Clean the equipment trolley and the measuring feet as required using a domestic cleaning
agent or a commercially-available disinfectant. Follow the manufacturer's instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for
cleaning. Such substances can damage the high-quality surfaces.
8. What do I do if...
there is nothing in the display?
– Is the scale switched on?
– Is the plug contact on the scale properly connected??
0. 0
does not appear before the weighing operation?
– Start the scale again. Do not change the load on the bed during switching on.
– Check whether the measuring feet are correctly positioned under the bed and that the
oscillating arms for the bed castor are freely suspended.
the scale goes off again after a few seconds?
– The mains unit is not connected. Connect the mains unit to a socket and start the scale
again.