Instrument Connections; Connection To Mains Power Supply; Power Supply Connection (Low Voltage Instruments); Power Supply Connection (High Voltage Instruments) - Hach ORBISPHERE 51x Manual Básico Del Usuario

Ocultar thumbs Ver también para ORBISPHERE 51x:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 89

4.4 Instrument connections

Figure 1 Connections - panel (left); wall/pipe (right)
1 Power cable
2 Ethernet cable gland
3 USB-A host connector
4 USB-B 4-pin connector
5 Input/Output 1 cable gland

4.5 Connection to mains power supply

4.5.1 Power supply connection (low voltage instruments)

For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 8-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
Figure 2 BINDER connector

4.5.2 Power supply connection (high voltage instruments)

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
10 English
6 K-M1100 Sensor connection
7 External pressure sensor connection
8 Input/Output 2 cable gland
9 TC Sensor connection
10 Keylock (wall/pipe mount only)
Pin Connections:
1. Power 10-30 VDC
2. Ground
3. Ground
4. Ground
5. Not used
6. Power 10-30 VDC
7. Power 10-30 VDC
8. Earth
D A N G E R
Figure 3 Wiring side view

Publicidad

Tabla de contenido
loading

Tabla de contenido