Specification
Details
Accuracy
Low range sensors
High range sensors
Limit of detection
Low range sensors
(LOD)
High range sensors
Response time
Low range sensors
(90%)
High range sensors
Display resolution
High and low range sensors
Calibration
Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use (air)
Calibration sample
Low range sensors: Standard 99.999% N
High range sensors: Standard 99.999% N
M1100 12 mm (PG
246 x 47 mm - weight 0.6 kg
13.5) sensor (L x W)
9.69 x 1.85 in. - weight 1.32 lbs
K1100 and
143.50 x 49 mm - weight 0.74 kg
M1100 28 mm
5.65 x 1.93 in. - weight 1.63 lbs
sensor (L x W)
Calibration device
Weight 0.7 kg
Section 2 Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Section 3 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

3.1 Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
± 0.8 ppb or 2%,
whichever is the greater
± 0.02 ppm or ± 3%,
whichever is the greater
0.6 ppb
0.015 ppm
< 30 seconds
< 50 seconds
0.1 ppb
(quality 50) or equivalent oxygen free gas
2
(quality 30) or equivalent oxygen free gas, air
2
N O T I C E
± 0.02 mbar or 20 ppm gas (at
atm pressure) or 2% of
reading, whichever is the
greater
± 0.5 mbar or 500 ppm gas (at
atm pressure) or 3% of
reading, whichever is the
greater
0.015 mbar or 15 ppm gas (at
atm pressure)
0.4 mbar or 400 ppm gas (at
atm pressure)
< 10 seconds
< 10 seconds
0.001 mbar or 1 ppm gas
English 5