Medtronic Guardian Link 3 Manual Del Usario página 19

Ocultar thumbs Ver también para Guardian Link 3:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 103
Warning:
You must adhere to Standard Precautions when handling or using
this device. All parts of the system should be considered potentially
infectious and are capable of transmitting blood-borne pathogens
between patients and healthcare professionals.
The transmitter must be disinfected after use on each patient. This
device may only be used for testing multiple patients when
Standard Precautions and Medtronic's disinfection procedures are
followed.
To clean and disinfect the transmitter, you need the following materials:
•  gloves
•  mild liquid soap
•  soft-bristled toddler toothbrush
•  8.25% bleach
•  two containers
•  clean, lint-free dry cloths
Use life
The transmitter can be cleaned up to 122 times or one year, whichever comes first.
Discard the transmitter at this point. If you continue to use the transmitter beyond 122
times or one year, the cleaning process may damage the device. Contact Medtronic
to order a new transmitter.
To clean and disinfect the transmitter:
 1
Wash your hands and put on gloves.
 2
Inspect the inside of the transmitter connector opening for any sign of body fluid.
For instructions on how to inspect the connector pins, see Inspecting the
transmitter connector pins, on page 8.
Caution:
The person inspecting the transmitter must have sufficient vision
that enables him or her to see small drops of body fluid or debris.
-13-

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido