Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R

4. SETTING

4.1 Basic adjustments

4.1.1 Access to mode "Operating Data"

When using the centrifuge, the following parameters can be set:
- Temperature indication ° C or ° F
- Acoustic signal turn on/off
- Keyboard sound turn on/off
- Volume pre-selection of sound signal
- Song selection of sound signal "end of run"
While the centrifuge is turned off, press simultaneously the keys "Time"(6) and "Door Open" (7) and turn on the
main switch of the centrifuge. Now release both keys and as a result a display test is executed for approx. 5
seconds. All indicators will appear at the same time (See figure 29).
ATTENTION:
Please notice that you must enter the program as described under point 4.1.1 to change the
adjustments of the points 4.1.2 - 4.1.7. After you have stored the settings you can change to
normal program mode again by switching off the centrifuge for a short while.
All changed settings must be confirmed by the key "Start"(9). The word "Store"(12) appears in the display "RPM |
RCF"(A-1) - Only then the pre-selections are valid!

4.1.2 Temperature indication

Proceed as described under point 4.1.1 to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" appears. Now select the letter "C" with the adjusting knob (1). As
a result, in the display "RPM | RCF" (A-1), the words "CELSI/temp" appear. If you press the key "RPM | RCF" (4),
the word "CELSI" flashes and you can change the display into Fahrenheit "FAREN", with the adjusting knob (1)
(See figure 31).
Figure. 29
Figure. 30
A-1
EN-14