Recall Of Stored Programs; Leaving Program Mode; Tarting And Stopping The Centrifuge; Starting The Centrifuge - OHAUS Frontier FC5714 Manual De Instrucciones

Ocultar thumbs Ver también para Frontier FC5714:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

TM
Frontier
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R
EN-12
For adaption of data press the key "Store" (12) (refer to figures 23 and 24) for approx. 1 second. If the program is
stored correctly, the word "StorE" appears in the display "RPM/RCF" (A-1). As a result, the word "programm"
(M10) disappears.
As soon as the key "Store" (12) is released, the word "programm xx" (M10) reappears – the (xx) stands for the
chosen program location.
If all program numbers are occupied you can take an old number that is not necessary anymore and just put in the
new parameters.

3.5.2 Recall of stored programs

To recall stored programs press the key "Prog/Setup" (11) (refer to figure 25) while the lid is already closed. Inside
the display "Time/Prog" (A-3), "programm –"(M10) appears. The desired program number can be pre-selected
with the adjusting knob (1).
In the respective displays the stored values for that program will appear.
If the wrong rotor is inside the centrifuge for the pre-selected program, in the display "RPM | RCF" (A-1), the word
"rotor" (M3) flashes. At the same time the word "FALSE" and the stored rotor number " xx" (M4) will flashing by
turns.
M3
M4
A-3
A-1
M10
11
1
12
Figure. 25

3.5.3 Leaving program mode

To leave the program mode just press the key "Prog/Setup" (11) (refer to figure 25). Then inside the display
"Time/Prog" the word "programm" appears.
Set the display to "programm--" (M10) with the adjusting knob (1).
3.6 Starting and stopping the centrifuge

3.6.1 Starting the centrifuge

You can start the centrifuge either with the "Start" key (9) (refer to figure 26) or the "Quick Spin" key (8).
By the "Start" key (9) you can start stored runs or runs with manually pre-selected parameters.
When the respective pre-selected running time has ended the centrifuge will stop automatically.
By the "Quick Spin" key (8) you can start runs, which will last just a few seconds.
By pressing the "Quick Spin" key (8) the centrifuge accelerates up to the pre-selected revolution.
In the display "Time/Prog" (A-3) the passed running time is indicated from the date of pressing the "Quick Spin"
key (8).
By releasing the "Quick Spin" key (8) the centrifuge stops and the running time is indicated until the opening of
the lid.