Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R
WARNING: Lifting Hazard. Single person lift could cause injury. Use a mechanical lifting
device or team lifting procedures when lifting or moving the equipment.
 Place the device on a stable, horizontal and non-resonant lab bench
1. Remove the front and back transport protection material.
2. Remove the plastic sleeve.
3. Observe a minimum distance of 30 cm to adjoining devices at the sides and
from the rear side to the wall.
4. Install the device in a well-ventilated location which is protected from direct
sunlight to prevent it from overheating.
 Connect the device
1. After installation, wait for four hours before switching the centrifuge on in order
to avoid damage to the compressor.
2. Check that the mains voltage and frequency match the requirements on the
device name plate(see rear side of the device) and then connect the device to the
power supply.
 Remove the transport protection material from the rotor chamber
1. Switch on the device at the mains power switch.
2. Open the centrifuge lid using the open button.
4. Remove the transport protection material.
5. Place the rotor vertically onto the motor shaft.
6.Turn the rotor nut using the rotor key clockwise until the rotor nut is tightened.
 The device is now ready to use
Retain the packaging and all transport protection material for shipping the
device at a later date.
8.3.3 Packing
Pack the centrifuge in reverse order.

8.3.4 Passing on the Device

When passing the equipment on to third parties, please make sure to also include