Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Medel CONNECT CARDIO MB10 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 109

Guidance and manufacturer's declaration - electromagnetic immunity

The MB10 is intended for use in the electromagnetic environment specified below. The customer or
the user of MB10 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for
transient/burst
power supply lines
IEC 61000-4-4
Surge IEC
±1 kV line(s) to line(s)
61000-4-5
±2 kV line(s) to earth
Voltage dips,
<5% U
short interruptions
(>95% dip in U
and voltage vari-
for 0.5 cycle
ations on power
40% U
supply input lines
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
100029_MEDEL CONNECT CARDIO MB10_USER MANUAL_REV.00_20.06.2016.indd 188
IEC 60601 test level
T
)
T
T
)
T
T
)
T
T
)
T
Compliance level
Electromagnetic environment
±6 kV contact
Floors should be wood, concrete
±8 kV air
or ceramic tile. If floor are covered
with synthetic material, the relative
humidity should be at least 30%.
±2kV for
Mains power quality should be
power supply lines
that of a typical commercial or
hospital environment.
±1 kV line(s) to line(s)
Mains power quality should be
±2 kV line(s) to earth
that of a typical commercial or
hospital environment.
Mains power quality should be
<5% U
T
that of a typical commercial or
(>95% dip in U
)
T
hospital environment. If the user
for 0.5 cycle
of the MB10 requires continued
40% U
operation during power mains
T
(60% dip in U
)
interruptions, it is recommen-
T
for 5 cycles
ded that the MB10 be powered
from an uninterruptible power
70% U
T
supply or a battery.
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
188
- guidance
23/06/2016 14:35:41

Publicidad

Tabla de contenido
loading

Tabla de contenido