Tunneling the Lead or Lead Extension
If using a 4x8 Surgical Lead or Infinion CX Lead, it is recommended to use the Long Tunneling Tool
(35 cm).
1. If not already assembled, attach the tunneling tool handle to the shaft by turning the locking
mechanism clockwise.
Tool Handle
Locking Mechanism
2. Mark the desired route of the tunnel.
3. Administer the appropriate local anesthetic along the tunneling path.
4. OPTIONAL. If necessary, bend the tool shaft to conform to the patient's body.
5. Make a small incision at the desired exit site.
6. Create a subcutaneous tunnel between the lead(s) incision and the IPG pocket site until the straw is
visible and accessible at the exit point.
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Shaft
Spectra IPG Entrada 2.0 19L
Instructions for Use
Implantable Pulse Generator
90970875-06 Rev A 7 of 285