Descargar Imprimir esta página

Specifications And Technical Data; Materials - Boston Scientific SC-1132 Instrucciones De Uso

Generador de impulsos implantable

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 17
Implantable Pulse Generator

Specifications and Technical Data

Parameter
Areas (Channels)
Amplitude
Rate
Width
Cycle
Ramp ON
Contacts
a. Only one Area is available if the rate is >130 pps.
b. Amplitude × Width ≤ 12.7 µC for all leads other than the 4x8 Surgical Lead; Amplitude × Width
≤ 10.6 μC for the 4x8 Surgical Lead.

Materials

Case
Header
Strain Relief
Size/Volume
Implantable Pulse Generator
90970875-06 Rev A
4 of 285
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Range
4
0 – 25.5 mA
2 – 1200 pps
a
20 – 1000 μsec
b
1 sec – 90 min, OFF
1 – 10 secs
1 – 32, case: +100% to -100%, OFF
Titanium
Epoxy
Silicone
55.0 mm x 46.0 mm x 10.8 mm / 21.2 cm
Spectra IPG Entrada 2.0 19L
Default
0 mA
40 pps
210 μsec
OFF
3 secs
1 – 32, case: OFF
(including header)
3
92070451 A.1
Page 8 of 290

Publicidad

loading

Este manual también es adecuado para:

Precision spectra sc-1132