Implantable Pulse Generator

Max Current Amplitude per Electrode Vs Impedance

(for all leads other than the 4x8 Surgical Lead)

Maximum Amplitude Based on Impedance and Pulse Width
Programmable up to 25.5 mA
25.50
Programmable up to 21.1 mA
20.40
Programmable up to 15.8 mA
15.30
Programmable up to 12.7 mA
10.20
5.10
0.00
300
400
Max Current Amplitude per Electrode Vs Impedance

(for the 4x8 Surgical Lead)

Maximum Amplitude Based on Impedance and Pulse Width
Programmable up to 25.5 mA
25.50
20.40
Programmable up to17.6 mA
15.30
Programmable up to 13.2 mA
Programmable up to 10.6 mA
10.20
5.10
0.00
300
400
Implantable Pulse Generator
90970875-06 Rev A
2 of 285
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
500
600
700
800
Impedance (Ω)
500
600
700
800
Impedance (Ω)
Spectra IPG Entrada 2.0 19L
900
1000
1100
1200
900
1000
1100
1200
Pulse Width:
PW = 20 us
20 µsec
PW = 100 us
100 µsec
PW = 200 us
200 µsec
PW = 300 us
300 µsec
PW = 400 us
400 µsec
600 µsec
PW = 600 us
800 µsec
PW = 800 us
1000 µsec
PW = 1000 us
Pulse Width:
PW = 20 us
20 µsec
PW = 100 us
100 µsec
PW = 200 us
200 µsec
PW = 300 us
300 µsec
PW = 400 us
400 µsec
PW = 600 us
600 µsec
800 µsec
PW = 800 us
1000 µsec
PW = 1000 us
92070451 A.1
Page 6 of 290