Omron SpotArm i-Q132 Manual De Instrucciones página 42

Ocultar thumbs Ver también para SpotArm i-Q132:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 202
HEM-1010-E_[D].book 42 ページ 2011年11月15日 火曜日 午後7時35分
4. Handling Errors and Problems
Problem
The date and time flash,
even though the date
and time have been set,
after the batteries are
replaced.
No display appears
when O/I button is
pressed.
Readings are not stored
in the memory.
42
Cause
The emergency deflation
switch was pressed.
Have the batteries been
removed?
The batteries are empty.
The batteries are
installed incorrectly.
The AC adapter was
disconnected from the
unit when using the AC
adapter only.
The AC adapter was
disconnected from the
power supply when using
the AC adapter only.
G (guest) was selected,
or no user ID was
selected.
Remedy
Set the date and time
again. (Refer to Chapter
2.3.)
Even if the power is
turned off, the clock will
also stop if you press the
emergency deflation
switch. Take care not to
press the emergency
deflation switch when
cleaning or moving the
unit.
Set the date and time
again. (Refer to Chapter
2.3.)
The date and time will be
reset if the batteries are
removed for more than
30 seconds. Replace the
batteries within 30
seconds.
Replace the batteries.
(Refer to Chapter 2.1.)
Insert the batteries
correctly.
(Refer to Chapter 2.1.)
Connect the AC adapter.
(Refer to Chapter 2.2.)
Connect the AC adapter
to the power supply.
(Refer to Chapter 2.2.)
Select the correct user
ID. (Refer to Chapter
3.3.)

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido