IRREGULAR HEARTBEAT SYMBOL (

When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol (
the display with the measurement values.
An irregular heartbeat rhythm is defined as
a rhythm that is more than 25% slower or
25% faster from the average rhythm
detected while the monitor is measuring the
systolic blood pressure and the diastolic
blood pressure.
If the irregular heartbeat symbol (
displays with your measurement results, we
recommend you consult your physician.
Follow the directions of your doctor.

MOVEMENT ERROR SYMBOL (

The Movement Error Symbol is displayed if you move your body during
the measurement. Please remove the arm cuff, and wait 2-3 minutes.
Take another measurement, remain still during inflation.

USER ID SYMBOL

The USER ID Symbol indicates USER A or USER B as selected by
the user to take a measurement or when using the memory function.
The USER ID Symbol is not displayed when using the Guest Mode to
take a measurement.