Answer = after a request from the host, the
Turbo-V 250 PCB controller will answer in one of the
following ways:
•
ACK
•
NACK
•
Message
When the Request is "A, "B", "C", "D", "F", the Turbo-V
250 PCB controller will Answer the ACK or NACK.
Converted ASCII
for CRC calculation
-
BFh
-
BEh
-
BDh
-
BCh
-
BBh
-
BAh
-
B9h
-
-
B7h
-
B6h
-
B5h
-
FAh
-
EBh
When the Request is "E, the Answer will contain the
complete set of the following parameters:
Bytes
0000=STOP
0001=WAITING INTERLOCK
LSB X
0010= STARTING
X
0011 = NORMAL OPERATION
X
0100= HIGH LOAD
X
0101= ""
1
0110 =FAILURE
0111=APPROACHING LOW SPEED
X
...
X
...
X
...
MSB X
...
2÷5
Cycle Time
Pump Life
6÷9
Pump Temperature
10÷11
12
Current*
13
Voltage*
Frequency
14÷17
18÷19
Cycle #
20
R1 Status
21
R2 Status
22
CRC
*The values for current and voltage are given as numbers,
scaled from 0 to 255, where 0 corresponds to 0 V and 255
to the full scale voltage (100 V) or current (2,5 A).
When the Request is "G", the Answer will contain a string
of 11 characters with the following parameters:
Bytes
1-2
Pump cycles number (integer
coded in 2 bytes)
3
Speed threshold
4-7
Run up time in seconds
(long coded in 4 bytes)
8
Deat time
(0 = NO 1 = YES)
9
Controller OFF when R2 OFF
(0 = NO 1=YES)
10
Soft Start Mode
(0 = NO 1=YES)
11
CRC
When the configuration parameters have to be changed,
send a Request string with the letter "H" and a series of
characters as shown below:
"H" + .... data .... + CRC
57