Indications and Adverse Effects
Indications
The AutoSonix™ hand instruments have application in abdominal, pediatric, gynecologic and other open and endoscopic procedures for the
transection, dissection and coagulation of tissue(s).
Adverse Effects
Certain parts and accessories to be used in conjunction with the AutoSonix™ System are intended for repeated use on different patients.
Such parts and accessories are clearly defined in this manual and MUST BE PROPERLY DECONTAMINATED AND STERILIZED BEFORE EACH USE
AS PER INSTRUCTIONS CONTAINED IN THIS MANUAL. FAILURE TO DO SO MAY LEAD TO ADVERSE EFFECTS SUCH AS, BUT NOT LIMITED TO,
INFECTIONS AND TRANSMISSION OF DISEASES SUCH AS HEPATITIS.
Device Specifications
Ultrasonic Generator
Equipment Type: Class 1 Type CF
Power Input: 110-130VAC 60Hz /200-240V AC 50Hz 250 VA
Operating Frequency: 55.5 kHz ± 500 Hz
Ground Leakage Current: 150 µ A maximum
Mode of Operation: Continuous Wave
Features:
• Main Power ON/OFF Switch (on rear panel)
• Power On Indicator (LED)
• Line Input Voltage Selector Switch (on rear panel)
• Ready/Standby Switch
• Amplitude Control Knob
• Output Amplitude Level Indicator (digital)
• Ultrasound On Indicator Light
• Mechanical Limit Indicator Light
• Electrical Fault Indicator Light
Operating Conditions: Temperature 55° F - 95° F (12.7° C - 35° C), Relative Humidity 20-90% (non-condensing)
Dimensions: 7 1/2" W x 19" D x 11" H (19 cm W x 48.3 cm D x 27.9 cm H)
Weight: 19 lbs. (8.6 Kg)
Transducer
Operating Frequency: 55.5 kHz ± 150 Hz
Power Capability: 80 Watts (max)
Operating Conditions: Temperature 55° F - 95° F (12.7° C - 35° C)
Dimensions: 5 1/2" L x 1 1/2" Dia. (14.0 cm L x 3.81 cm Dia.)
Unpacking Instructions
Carefully inspect shipping container before opening. After visual inspection of container, carefully unpack all components and place on
a table or cart. Visually inspect all components for obvious shipping damage. Retain the shipping container and immediately notify the
shipping carrier of any damage.
The following components are included:
(1) Generator Box
(1) Footswitch
CAUTION: Use only genuine replacement parts manufactured for United States Surgical. Use of parts furnished by other sources may
result in danger or failure and may void any warranty that may apply.
11