Tecno-gaz Hydra EVO Instrucciones De Uso página 27

Ocultar thumbs Ver también para Hydra EVO:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
06
The instruments to be sterilised must be previously cleaned, free from residual materials,
blood, deposits, etc. The above mentioned rests can damage the materials to be sterilised
or the autoclave components; moreover, they do not ensure a perfect sterilisation.
1. Wash the instruments after using in cold water to remove any rest.
2. Insert the instruments in the ultrasound machine.
3. Rinse the instruments with clean water with care.
4. Dry the instruments to avoid that not demineralized water remains on them, this causing
oxidation or instruments spots due to the calcareous rests.
5. Wash, rinse and dry also the supplied trays.
6. On the same tray only instruments of the same material must be placed.
7. The small mirrors must be positioned with glass downwards.
8. For a better sterilisation open the instruments such as forceps, shears or other
composed instruments.
9. Do not stack the instruments on the trays. The instruments must be sterilised separately.
An overload could compromise the sterilisation.
10. If empty containers are sterilised to avoid a water accumulation, it is necessary to place
them upside down.
11. Sterilise the instruments only on the trays inserted in the supplied tray - holder; there
must be a certain distance between the trays so that vapour circulates during the
sterilisation phase, this easing drying.
07
CYCLES
121°C
(not wrapped)
134°C
(not wrapped)
The choice of the sterilisation cycle must be based on the data supplied by the
INTRODUCTION FOR STERILISATION
STERILISATION TABLE
STERILISATION
TIME
[minutes]
22,5
8
producer of the object you want to sterilis
DRYING
WORKING
TIME
PRESSURE
[minutes]
[Bar Relative]
5
5
2,25 ± 0,2
ENGLISH
1,3 ± 0,2
9

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hydra evo plus2066-s-22066l-s-2

Tabla de contenido