Error Message
Energy < 80%*
Energy > 120%*
Energy > 200%
Maintenance due
MPS charge error
MPS over temp
MPS overload
*Minor error. Schedule service but ok to continue using laser.
Cook Medical © 2012
Possible Cause
• Optical component damage
• Incorrect energy calibration
• Incorrect energy calibration
• Preventative maintenance is
due
• Main power supply
malfunction
• Main power supply
malfunction
• Main power supply
malfunction
Return to Table of Contents
Maintenance
Solution
• If the energy error persists,
call for service*
• If the energy error persists,
call for service*
• Call for service
• Call for service*
• Call for service
• Call for service
• Call for service
Chapter 6
1-36