7-4 Replacement of the Bulb (Optic Motor ONLY)
1) Loosen Motor Housing and detach from the motor, and remove the old light bulb using the small pin, like the
precision screwdriver. (Fig. 22)
2) Align and insert the connector of the new bulb into the Socket Hole securely. Screw the motor housing securely to
the motor.
Optional Bulb: TA Bulb (Pack of 3)
*
Motor Housing
CAUTION
• Make sure the power 'OFF' (remove the AC Electrical cord)
• Do not touch the glass of new bulb.
7-5 Maintenance of the Control Unit and Foot Control
If blood or saline solution has stained the Control Unit or Foot Control, remove the AC Electrical Cord; wipe Control Unit
or Foot Control with a damp cloth, then with an alcohol-absorbed cloth.
7-6 Maintenance of the Handpiece Attachment
For detail, refer to the attached handpiece Operation Manual.
8. Sterilization
• Autoclave sterilization is recommended.
• Autoclave sterilization is required for the first time use and after each patient as noted below.
The following items can be autoclaved.
• Implant Handpiece
• Micromotor with Motor Cord (Including the Motor Housing)
• TA Bulb
• Handpiece Stand
• Internal Irrigation Nozzle
• Tube Holder
• Nozzle Holder
• Protection Plug
• Calibration Bur
CAUTION
Only Items Specified above can be autoclaved
Order No. Y900132
A
Bulb
15
Bulb Socket
Connector
Bulb
View from A
Fig. 22