Optional Accessories; Connector; Nozzle Holder 1 (*2); Internal Irrigation Nozzle - NSK Surgic XT Plus Operación Manual

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Implant Handpiece can be cleaned and disinfected with a Thermo-Disinfector.
[Autoclaving]
1) Remove blood and debris from the handpiece.
2) Clean inside the handpiece; by using the spray lubricant (refer to "7-6. Maintenance of the Handpiece Attachment").
• Do not attempt to spray lubricant into the Micromotor.
• Attach the Protection Plug to the Micromotor. (Fig. 23)
3) Place those in autoclave pouch (not included in the package) and
seal it.
4) Autoclave up to max. 135˚C. (Autoclave for 20 min. at 121˚C, or 15
min. at 132˚C.)
5) Keep the handpiece in the autoclave pouch to keep it clean until you
use it.
Sterilization at 121 ºC for more than 15 minutes is recommended
*
by EN13060 or EN ISO17665-1.
Caution for Autoclaving
• Clean and lubricate the handpiece before autoclaving. Autoclaving a handpiece stained with blood or debris
could cause damage to the handpiece.
• Do not lubricate the Micromotor.
• Do not disconnect the Motor Cord from the Micromotor.
• The Irrigation tube is a single use disposable type and cannot be autoclaved.
• Do not use Protection Plug without the mounting O-Ring. It may cause a malfunction. If O-Ring had
damaged, replace it immediately.
* O-Ring (Protection Plug): Order code 0312457102
• You can hang Protection Plug on the Hunger Post to prevent loosing it.
• Do not hung the Protection Plug with Micromotor. It may drop the motor and cause damage.

9. Optional Accessories

Part No.
C823752
Y900113
20000396 (for X-SG20L)
20000357 (for SG20)
10000324
Y900083
Z182100

Description

Y-Connector
Irrigation tube (Pack of 5)
Nozzle Holder

Internal Irrigation Nozzle

Tube Holder

PANA SPRAY Plus

Protection Plug

Remarks
Used for branching the internal and external coolant
irrigation.
For replacement tube use.
Supplied as standard accessory items.
Supplied as standard accessory.
Supplied as standard accessory for Irrigation Nozzle.
For High & Low speed handpieces.
16
Micromotor
Fig. 23