Thank you for purchasing the NSK Surgic XT Plus surgical unit.
We recommend that prior to use, you carefully read this document regarding instructions for use, handling method, or
maintenance check so that you can carry on using the unit in the future. In addition, keep this operation manual in a
place where a user can refer to it at any given time.
Classification of equipment
• Type of protection against electric shock :
– Class l equipment
• Degree of protection against electric shock :
– Type BF applied part
• Method of sterilization or disinfection recommended by the manufacture :
– See 8. Sterilization
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529 :
– Foot Control: IPX8 (Protected against the effects of continuous immersion in water)
• Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide :
– Motor, Foot Control: Category AP Equipment
•Mode of operation :
– Intermittent operation
Cautions for handling and operation
Read these safety cautions thoroughly before use and operate the product properly.
These indicators are to allow you to use the product safely, prevent danger and harm to you and others. These are
classified by degree of danger, damage and seriousness. All indicators concern safety, be sure to follow them.
Classification
WA R N I N G
Explains an instruction where personal injury or physical damage may occur.
C A U T I O N
Explains an instruction where minor to medium injury or physical damage may occur.
N O T I C E
Explains an instruction that should be observed for safety reasons.
1. Safety precautions prior to use
Read Handpiece's Operation manual carefully before use.
Surgic XT Plus is intended for use in dental oral surgery and surgical procedures by qualified personnel.
The system may present a possibility of malfunction when used in the presence of an electromagnetic interference
wave. Do not install the system in the vicinity of the device which emits magnetic waves. Turn off the power switch of
the Control Unit of this system when an ultrasonic oscillation device or an electrode knife is located in the vicinity is
used.
• Surgic XT Plus needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information.
• Portable and mobile RF communications equipment can affect Surgic XT Plus. Do not use RF equipment outskirts for
the product.
• The use of accessories, Motors and cables other than those specified, with the exception of Motors and cables sold
by the manufacturer of Surgic XT Plus as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of the Control Unit.
Degree of Danger or Damage and Seriousness
Intended to Use
WARNING
CAUTION