8. CALIBRATION
Turn on the scale. Keep pressed
Pn
Display will show "
" briefly PIN is activated. Press the PIN number to get into the parameter menu.
•
The default PIN number is "0000"
•
Press numeric key 0 four times, display will be shown Pn - - - -
•
Press
key to confirm, it will enter into the parameter settings and will to show "
Then press the calibration switch to access to calibration.
•
Press
key to confirm for calibration, display will be show "
•
Remove the goods from the platform and make sure STABLE indication has been enable.
•
Press
key to confirm.
Display will be show ,CAP LoAdXX.
•
•
If necessary change the calibration weight value by pressing
•
Press
key to confirm when display gets stable, display will be show PASS.
Finished calibration and display will come to FO CAL.
•
Remove calibration weigjt from the platform.
•
Press
key for to go back to the weighing mode.
9. RS 232 OUTPUT
1. SPECIFICATIONS:
RS-232 output of weighing data
ASCII code
1200~115200 Baud
8 data bits
No Parity
2. RS-232 (9PIN D TYPE CONNECTOR)
Pin 2
Pin 3
Pin 5
key during the self test.
RXD
Input
TXD
Output
GND
-
F0 CAL
UNLOAD
"
key and place the callibration weight on the platform.
Receiving data
Transmission data
Signal ground
USER MANUAL XTA
EN
".