5.2 CALIBRATION / ADJUSTMENT

REQUEST
The machine only can be Calibration / Adjustment in the following cases:
Nothing on weighing pan, was tare, weighing signal stable
Machine will have ERROR SIGNAL if does not have above premise.
If all premise available, the machine will display required weight mass value for calibration.
INSTANCE
EXTERNAL SPAN CALIBRATION INSTANCE
KEY (ORDER)
Press and Hold [CAL] Key
EXTERNAL LINEARITY CALIBRATION INSTANCE
KEY (ORDER)
Short Press [TARE] Key
Press and Hold [CAL] Key for
5 seconds
Release [CAL] Key,
Press and Hold [MENU] Key
for 5 seconds
NOTE: The grey color words explain the signal's meaning which flash on the window.
78
STEP EXPLANATION
1. Machine tare the weight
2. Display span calibration signal Flash the required weight mass value:
100g
3. Put required weight mass on pan. After 5 seconds, display the weight
of it
4. Take the weight mass away (Span calibration finished)
STEP EXPLANATION
1. Machine tare
2. Display Span CAL signal Flash 100g signal
3. Display Linearity CAL signal Flash 100g signal
4. Put required weight mass on pan Display 100g after 5 seconds
5. Take weight mass away Linearity calibration step Flash 50g signal
6. Put required weight mass on pan Display 50g after 5 seconds
7. Take weight mass away (Linearity calibration finished)
USER'S MANUAL ABS
LCD SCREEN DISPLAY
0.0000 g
100.0000 g
100.0000 g
0.0000 g
LCD SCREEN DISPLAY
0.0000 g
100.0000 g
100.0000 g
100.0000 g
50.0000 g
50.0000 g
0.0000 g
EN
EN