MC2000 Moisture Analyzer Series

3.3.2 Standby Mode

When the instrument is installed batteries and the display is off, the device is
in Standby mode.
Start
(Short press) Turns on the display, self-check and enters category
mode. The category code is showed on the display "C##".
Stop
(Long press) Re-enters standby mode. (off)
3.3.3 Category Mode
(Short press) Changes modes to load category, show "C##" on
Category
display.
Up
(Short press) Select the category code.
Down
(Short press) Select the category code.
Category
(Short press) Load category parameter and exit to measurement
mode.

3.3.4 Measurement Mode

(Short press) Changes modes to measurement, show "0.0%" on
Category
display.
Up
(Short press) Shows sample weight value (after moisture
measurement)
Up
(Press twice) Shows the density of sample (after moisture
measurement)
Down
(Short press) Shows the temperature value of sample (after
moisture measurement)
Enter
(Short press) Shows mean moisture value and delete the
previous measurement result (after at least two measurements)
3.3.5 Adjust Mode
Adjust
(Long press) Enters adjust mode.
Up
(Short press) Increase adjust value.
Down
(Short press) Decrease adjust value.
Category
(Short press) Save and enter measurement mode.
Enter
(Short press) Save and enter measurement mode.
4. ADJUST

4.1 Weight Calibration

Before using the analyzer for the first time, we recommend to calibrate it.
1.
Put the instrument on a level surface, remove top cover and empty
the instrument.
Press "Category" button, changes mode to measurement
2.
Long press the "Up" button until "0.0" is flashing.
3.
4.
Place a 200g calibration mass on the sensor (top of red cone).
Remove the mass when "200" is shown on the display.
5.
Press the "Category" button to exit.