FLAEM Wi.Neb Go Manual De Instrucciones página 23

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
CLEANING, SANITIZATION AND DISINFECTION
Turn off the unit before any cleaning and disconnect the power cord form the mains.
CLEANING -UNIT (A) AND OUTER SURFACE OF THE TUBE (B)
Only use a cloth dampened with antibacterial detergent (non-abrasive and free of solvents of any kind).
washer with a hot cycle.
method C: Sanitise accessories G1-G2-G3-G4-H-I-L-M-F by immersing them in a solution of 50% water and 50% white vinegar, then rinse
thoroughly under potable hot water (approximately 40°C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry them with a jet of hot
air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the methods described below.
method A: Accessories G1-G2-G3-G4-H-I-L-M-F can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite) specific for disinfecting, which is available
in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable water and disinfectant, according to the propor-
tions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the parts. Leave the parts immersed
for the amount of time indicated on the packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B: Sanitise the accessories G1-G2-G3-G4-H-I-L-M-F by boiling them in water for 10 minutes; use demineralised or distilled water
to prevent calcium deposits.
method C: Sanitise the accessories G1-G2-G3-G4-H-I-L-M-F with a hot steam steriliser for baby-bottle (not the microwave type). Perform
the process faithfully following the instructions of the steriliser. To ensure that the disinfection is effective, choose a steriliser with an
operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry them with a jet of
hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away from dust.
NEBULIZER AND ACCESSORIES
Open the nebuliser by turning the upper part (G1) anticlockwise, remove
the nozzle (G2) from the upper part (G3) by pressing as shown by the 2
arrows in the "Connection diagram" in point G.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories,
choosing one of the methods described below.
method A: Sanitise accessories G1-G2-G3-G4-H-I-L-M-F under potable
hot water (approximately 40°C) with a gentle, non abrasive dish deter-
gent.
method B: Sanitise accessories G1-G2-G3-G4-H-I-L-M-F in the dish-
ENGLISH
- 21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

P0905em

Tabla de contenido