The device is configured at the factory according to the following parameters:
Programme
Very
D1
powerful
Powerful
D2
Average
D3
power
Low
D4
power
c) Timer
The device has a function that indicates the duration of use of the device. The total duration is displayed on
screen (Fig. 5, item 7) under the logo
The shortest duration displayed is one minute.
d) Software version
The software version is displayed at the bottom of the screen (Fig. 5, item 6).
e) Storing settings
To store the modified parameters and use the device, hold down button
screen shown in Fig. 6 reappears.
VIII – SAFETY
The device is fitted with a system that can detect malfunctions that may occur.
When an internal operating error occurs at start-up, symbol 1 (Chap XVII - Symbols) lights up (Fig. 7, item 1) and
4 beeps sound.
When an internal operating error occurs during operation, symbol 1 (Chap XVII - Symbols) lights up (Fig. 8, item
1) and 4 beeps sound. We recommend that you shut down the device with the mains isolating switch (Fig. 2,
item 5) then switch the device back on.
When the ultrasound function is activated and the handpiece is not connected or is not properly connected,
symbol 2 (Chap XVII - Symbols) lights up (Fig. 9, item 1) and 4 beeps sound. We recommend that you connect the
handpiece or check that it is correctly connected to the device.
IX – USING THE DEVICE
Important:
- Do not disconnect the handpiece cable when the device is on and the pedal is pressed down.
- Do not tighten or loosen the tips when the handpiece is activated.
- All accessories used shall be previously cleaned, disinfected and sterilized.
- For your safety and the safety of your patient, the device shall only be used with accessories supplied or
recommended by SATELEC.
- Before and after each use, check the device and its accessories to make sure there are no problems.
- If necessary, do not use the device and replace any faulty parts.
Operate the device as follows:
- Make sure that the device is correctly plugged in and corrected insulated.
- Set the mains isolating switch (Fig.2, item 5) to position I (ON).
- Hang the bottle or pouch containing physiological saline solution or sterile water from the bracket.
- Lift the lid of the irrigating pump.
- Horizontally insert an irrigation line cassette into the purpose-designed housing.
- Lower the lid.
- Connect the irrigation line perforator to the bottle or pouch containing physiological saline solution or sterile
water.
- Set the required parameters (irrigation flow rate, programme), see chapter VII.
Piezotome Solo V6 V2
Irrigation flow
Main
rates
functions
60ml/min
Osteotomy,
60ml/min
osteoplasty
60ml/min
Soft
60ml/min
removal
(Fig.5, item 3).
tissue
for a while (Fig. 3, item 4). The
10