MAINTENANCE AND TROUBLESHOOTING
The machine does not need any particular maintenance apart from the periodic
cleaning of the heating unit which is necessary to avoid the deposit of substances
inside the seat of the biological indicators.
For cleaning of the incubator is sufficient to refer to the most elementary rules of
hygiene health provided in such circumstances.
• First turn off the machine and pull the plug from the outlet
• If the machine has just been used, waiting for cooling.
• Use a sponge slightly moistened with soapy water and wipe dry with a dry and
clean
• Do not use detergents which could alter in irreversibile way the appearance of
plastic parts of the machine.
• Do not spray directly with nebulizer on the machine, risking damage to the
electrical equipment inside.
DO NOT DO MAINTENANCE OR INSIDE REPAIR TO THE DEVICE.
DISMANTLE THE DEVICE MEANS BREAKING IT.
Defect
Pushing START button
the incubator does not
work
The LED is turned on
during the test
After about 15 minutes
of operating time the
led switch on again
Cause
Intervention
There is no power
Control plug fit correctly in autoclave socket
supply
Resistance is
Contact maker for technical assistance
burned
Push start button; led switch off,remove tubes
(biological testing is to be considered unsuccessful)
and leave it on for about 15 minutes.
If led don't switch on automatically, means that the
Electrical energy
previous stopping was due to a temporary lack of
interruption
electrical energy.
Otherwise if device switch on, means that the
resistance is short‐circuit and security locks.
Contact maker for technical assistance
The heating unit
Try to repeat the cycle.
not arrived in
In the case of persistent problem contact maker for
temperature
technical assistance
Inside failure
15
ENGLISH