Operating States - ProMinent DULCOMETER Pt 1000 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Function Description
5.1
During normal operation, the meter can be in seven different states. These, as
well as the reaction of the meter when operating the keys "UP" or "DOWN", are
described below.
Operating state
Measure pH
Measure redox
Measure Pt 1000
SIM pH
SIM redox
30

Operating states

Markers
Operation
"pH"
The meter measures the signal of the
marker
connected pH probe and indicates the
lights up
corresponding temperature-corrected pH
value on the display. The associated LCD
marker "pH" is activated. When the
measuring range limits are exceeded, both
the "ERR" and "pH" markers will flash.
"mV"
The meter measures the signals of the
marker
connected redox probe and indicates the
lights up
measured value on the display. The
associated LCD marker "mV" is activated.
When the measuring range limits are
exceeded, both the "ERR" and "mV" markers
will flash.
"°C"
The meter measures the signals of the
marker
connected Pt 1000 probe and indicates the
lights up
measured value on the display. The
associated LCD marker "°C" is activated.
When the measuring range limits are
exceeded, both the "ERR" and "°C" markers
will flash.
"SIM"
The meter outputs the pH value indicated on
marker
the display as the respective mV signal on the
and "pH"
right-hand connection socket, if no Pt 1000
marker
electrode is connected. By operating the keys
light up
"UP" or "DOWN" the output can be changed
in the respective valid range with the
respective valid display resolution.
"SIM"
The meter outputs the redox potential
marker
indicated on the display as the respective
and "mV"
mV signal on the right-hand connection
marker
socket, if no Pt 1000 electrode is connected
light up
there. By operating the keys "UP" or "DOWN"
the output can be changed in the respective
valid range with the respective valid display
resolution.
®
ProMinent

Publicidad

Tabla de contenido
loading

Tabla de contenido