Technical Specifications; Storage Conditions - Halyard ON-Q Instrucciones De Uso

Ocultar thumbs Ver también para ON-Q:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
Figure 10
Note: Cover will not close if dialed between numbers (Figure 10).
Changing the Flow Rate During an Infusion
1. Insert the dial key into the dial of the SELECT-A-FLOW*
device.
2. Turn the dial key until the new flow rate is selected. Make
sure the selected flow rate setting is within the window
and aligned with the ml/hr x mark on the face of the
SELECT-A-FLOW* device (Figure 7, 8).
• The tactile feel will allow the user to ensure selected
flow rate is set.
3. Remove the dial key from the device and put in a safe place
for later use.
During the Infusion:
• A change in appearance and size of the pump may not be
evident during the first 24 hours after start of infusion.
• As medication is delivered, the pump will gradually
become smaller.
• Make sure:
• Clamp is open
• There are no kinks in the tubing
• Filter vent is not taped or covered
• Heat, ice or cold therapy is placed away from the flow
controller
End of Infusion:
• Infusion is complete when pump is no longer inflated.
• Close clamp, disconnect and dispose of the pump
according to your institution's protocol.
Note: If pump did not perform as expected do not discard.
Contact Halyard Health for product return instructions at:
HalyardIrvineProductComplaint@hyh.com or 1-800-448-3569.
8

Technical Specifications

Delivery Accuracy: When filled to the labeled volume,
SELECT-A-FLOW* device delivery accuracy is ±20% of the
labeled rates when infusion is started 0-8 hours after fill and
delivering normal saline as a diluent at 22°C/72°F.
Typical Flow Curve
The flow rate may be higher or lower at the beginning and
end of the infusion (Figure 11).
Figure 11
Typical Flow Curve
Percent Delivery Time
Note:
Latex is not in fluid pathway or in contact with human. Refer to
Latex Sensitivity Technical Bulletin at halyardhealth.com.

Storage Conditions

Store under general warehouse conditions. Protect from light
sources and heat. Keep dry.
Rx only
= CAUTION: Federal (U.S.A.) law restricts this device
to sale by or on the order of a physician.
Additional U.S. and Foreign Patents may be issued and/or
pending.
*Registered Trademark or Trademark of Halyard Health, Inc. or its
affiliates. © 2015 HYH. All rights reserved.
For more information, please call
+1.949.923.2400 • 1.800.448.3569 (English only)
or visit www.halyardhealth.com for the latest product
information and Technical Bulletins.
To order additional Instructions for Use or
Patient Guidelines please email or call:
HalyardIrvine.ProductInquiry@hyh.com or
+1.949.923.2400
Nominal
100%

Publicidad

Tabla de contenido
loading

Tabla de contenido