About This Sensor; Safety - ProMinent DULCOTEST OZE 3-mA-2 ppm Instrucciones De Servicio

Ocultar thumbs Ver también para DULCOTEST OZE 3-mA-2 ppm:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1
The ozone sensor OZE is a diaphragm-covered, amperometric two-electrode
sensor. The sensor provides a standard signal of 4-20 mA via a two-wire
interface. This signal is largely independent of the flow rate. The sensor is
supplied with voltage via the two-wire interface.
In connection with the measuring and control unit DULCOMETER
Ozone and the metering pumps of ProMinent, the ozone concentration can
be optimally adjusted. Typical applications include the treatment of
surfactant-free freshwater such as swimming pool water, drinking water or
industrial and process water.
2
Proper use
The sensor may only be used to determine and control the ozone
concentration (O
The sensor may solely be used in swimming pool water, drinking water or in
waters of similar quality!
All other uses or modifications are prohibited!
The sensor is not suitable to check the absence of ozone!
The sensor may not be used in water with a high gas percentage!
The sensor may not be used in surfactant-containing water!
The sensor may not be used outside of the stated operating conditions!
The sensor is no safety component part!
CAUTION
• The sensor may only be used in ProMinent in-line probes of the type
• A free drain must exist at the outlet of the in-line probe.
• The voltage supply to the measuring device and thus the sensor may
ProMinent
®

About this sensor

Safety

)!
3
DLG III or DGM. Only by this the flow rate parameters are kept! (see
Chapter 15 "Technical data")
not be interrupted.
After voltage interruptions of more than 2 hours, the sensor must be
run-in and calibrated again (see Chapter 7.1 "Running-in period" and
Chapter 7.2 "Calibration").
About this sensor / Safety
®