Problem
Camera otherwise behaves
abnormally.
4.5
Changing the LED module in the reflected-light fluorescence illuminator
ATTENTION
The LED module in the reflected-light fluorescence illuminator has to be changed in any case
by an authorized service technician.
The Primo Star iLED microscope with reflected-light fluorescence illuminator is provided with
an interlock function to switch off the installed LED of the reflected-light illuminator as soon as
the reflected-light illuminator will be turned with regard to the stand or detached or if the lid
of the reflected-light illuminator will be removed.
Fig. 45
Lid of reflected-light illuminator
54
Care and Troubleshooting
Cause
The camera may have been brought
into a non-intended state.
• Disconnect the reflected-light illuminator from
• Detach the tube (see also section 3.7.1). To this
• Loosen the three fastening screws (Fig. 45/1) of
• Disconnect the plug-in connector of the LED
• Unscrew the two fastening screws of the LED
• Remove the LED module (Fig. 46/4).
• Insert the new LED module. Because of limited
415500-7344-008
Primo Star / Primo Star iLED
Troubleshooting
Restore all default camera settings via
the OSD menu or by pressing the
"Snap" and "AWB / OK" buttons for
two seconds. The camera will then
restart and should be in a normal state.
the power supply.
end, loosen the knurled screw, turn the tube
counterclockwise by approximately 90° and lift
it out of the dovetail mount.
the lid of the reflected-light illuminator (2.5 mm
Allan key).
module connection cable (Fig. 46/3) on the
circuit board. To this end, press the safety hook
against the connector to release it and pull out
the connector.
holder (Fig. 46/1) completely. Remove holder
(Fig. 46/2) and screws.
space, the flat side of the LED module must
point to the circuit board.
11/2013