Primo Star / Primo Star iLED
3.7

Converting the microscope

Unplug the plug-in power unit from
line power before converting the
microscope.
3.7.1

Changing the tube

• Loosen the clamping screw (Fig. 37/2), turn the
installed
tube
(Fig. 37/1)
approximately 90° (Fig. 37/A) and detach it on
the right side upward (Fig. 37/B).
For reasons of space, the tube may
also be clamped by the Allen set screw
supplied with the tube.
• Insert the tube to be installed, with the
eyepieces pointing to the right and its dovetail
mount slightly inclined, under the two holding
elements (Fig. 37/4) into the stand.
• Then, place the tube horizontally onto the
stand. In doing so, the groove on the underside
of the tube must be located above the third
holding element (Fig. 37/3) of the stand.
• Turn
the
tube
counterclockwise
(eyepieces pointing frontward), align it to the
stand
and
tighten
(Fig. 37/2).
For
space-saving
microscope (e.g. in a cabinet) the tube
may also be rotated by 180° to the
back.
3.7.2

Inserting color filters

• Move the condenser carrier fully up by turning
the knurled knob for vertical adjustment
(Fig. 14/22).
• Unscrew the cover cap (Fig. 38/3) from the
luminous-field diaphragm (Fig. 38/1).
• Put the desired filter – yellow, green or blue –
(Fig. 38/2) onto the mounting surface of the
luminous-field diaphragm and screw the cover
cap back into place.
11/2013
Start-Up and Operation
clockwise
by
by
45°
the
clamping
screw
storage
of
the
415500-7344-008
Fig. 37
Changing the tube
Fig. 38
Inserting a color filter