Configuring The Integrated Hd Ip Camera In The Binocular Tube - Zeiss Primo Star Instrucciones De Manejo

Ocultar thumbs Ver también para Primo Star:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 177
Primo Star / Primo Star iLED
3.5.4

Configuring the integrated HD IP camera in the binocular tube

• Press the Menu button on the tube's control panel.
The OSD menu for adjusting the settings will appear.
The following menus are available for configuring the camera:
Fig. 24
Color menu
AWB
The purpose of white balance is to calibrate the camera to the color temperature of the light
source used. The AWB function of the integrated camera supports the Continuous / Push button
modes.
In Continuous mode the white balance is carried out continuously. In certain specimens this may
result in incorrect colors.
To ensure that the white balance applies to the light source and not the main color of the
specimen, the push button mode can be set. Here the white balance is carried out only once if the
AWB / OK button is pressed outside the OSD menu.
Color temp.
This creates a white balance bias towards yellow (left) or blue (right). If AWB (see above) is set to
push button, the effect of this regulator is only visible after a white balance has been carried out
with the AWB button (outside the OSD menu).
Saturation
The camera's images are in sRGB color space with an 8 bit color depth per channel. The color
saturation per RGB value can be adjusted.
Extreme left:
Extreme right:
Gamma
The gamma correction of the camera can be adjusted. A choice of Gamma 1 and 0.45 is available.
11/2013
Start-Up and Operation
Gray scales
Maximum color saturation
415500-7344-008
Carl Zeiss
35

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Primo star iled

Tabla de contenido