1. Press and hold the
displayed.
2. The default unit is gram (g). To change the unit, press
to advance to the next unit.
3. Press
Set Custom to On in the Unit menu to enable and define the Custom Unit.
The Custom Unit is defined by entering three parameters, Factor, Exponent and LSD (least significant digit).
Define the Custom Unit as follows:
1. Determine how many custom units there are in 1 gram.
2. Convert the value to scientific notation, e.g. m x 10n.
3. Enter the value of m as the Factor setting.
4. Enter the value of n as the Exponent setting.
5. Enter the amount that the Custom Unit steps by as the LSD setting.
Enter the Factor and the Exponent and LSD.