Custom Unit = Conversion Factor x Grams.
The LSD is the value by which the displayed weight is incremented or decremented.
The conversion factor is used by the balance to convert grams to the custom weighing unit and is defined by
entering a factor and an exponent. The factor is a value between 0.1000000 and 1.999999 inclusive.
For example: One cup of chemical
The exponent moves the decimal point of the factor to the right for positive values or to the left for negative values.
For example: One cup of chemical
The LSD is the value by which weight is incremented or decremented.
LSD
Result
1
Display counts by 1
5
Display counts by 5
10
Display counts by 10
...
...
For example, if the displayed digital is 0.56 for one cup of chemical, the LSD should be set 100.
Enter this sub-menu to customize RS232 standard settings. Data may be output to either a printer or a PC.
Set the baud rate (bits per second).
1200
2400
4800
19200 = 19200 bps
38400 = 38400 bps
Set the data bits, stop bit, and parity.
8-NO-2
7-EVEN-1 = 7 data bits, even parity, stop bit 1
7-EVEN-2 = 7 data bits, even parity, stop bit 2
7-NO-1
GlobalScientificSupply
www.
0.5643834×1 g, the factor should be set 0.5643834.
10 g, the exponent should be set 2.
= 1200 bps
= 2400 bps
= 4800 bps
= 8 data bits, no parity, stop bit 2
= 7 data bits, no parity, stop bit 1