Air Liquide Respireo SOFT Facial Manual De Uso página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 27
2.8. Do not use the mask if there are signs of vomiting or nausea.
2.9. Keep away from light.
2.10. The mask should be cleaned and/or disinfected (see
illustrations in section F).
2.11. Before use, check the integrity of the mask. If damaged during
transport, notify your dealer.
2.12. In case of visible deterioration (cracks, tears, etc.), it is
necessary to eliminate and replace the damaged mask
component.
2.13. Do not leave the components of the mask unattended, some
of these may be swallowed by children.
2.14. It is recommended to wash the mask before the first use.
2.15. Do not disassemble the anti-asphyxia fitting (c) from the
mask body (b) under any circumstances, even when
cleaning the mask.
3 - MASK COMPONENTS (Section A)
a) cushion; b) mask body; c) anti-asphyxia fitting; d) swivel fitting;
e) headgear; f) buckle.
4 - MASK USE
To use the mask correctly, see the illustrations in Sections B - C -
D - E - G.
5 - REGENERATION OF THE MASK FOR REUSE ON DIFFERENT
PATIENTS
Notes:
- The mask should always be regenerated before being used on a
new patient.
- The headgear (e) must be replaced.
The mask can be regenerated for up to 20 cycles by one of the
following processes:
● thermal disinfection at a high level;
● chemical disinfection at a high level;
ENGLISH
Detailed mask regeneration instructions can be found on the Internet
at www.device.airliquidehealthcare.com or from your retailer.
6 - TROUBLESHOOTING
PROBLEMS
• No air comes out
of the exhalation
orifices.
• Difficulty
breathing.
• Noisy mask.
• Air leaks
between the
cushion and the
face.
21
POSSIBILE CAUSES SOLUTIONS
• Flow generator
• Turn on the flow
turned off or not
generator;
working.
• Adjust the flow
generator.
• Patient circuit not
• Connect the patient
inserted correctly.
circuit to the flow
generator and the mask.
• Vent ports are or
• Clean the orifices using
partially obstructed
a soft bristle brush;
• Disassemble the mask
• Mask is not properly
and then reassemble it
assembled
as described in the user
manual.
• Reposition the mask on
• Mask is not correctly
your face and adjust the
positioned
headgear as described in
the user manual.
• The mask is not
• Reposition the mask
properly positioned on
and adjust the headgear
the face.
as described in the user
• The straps of the
manual.
headgear are not
properly adjusted (too
tight or too loose).
• The size of the mask
• Contact your doctor or
may not be correct.
the technician to find the
correct size