TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R

3.5 Program

3.5.1 Storage of programs

You can store up to 99 runs with all relevant parameters, including the used rotors. You can use any free program
number and call it up again.
Put the needed rotor into the centrifuge. By pressing the key "Prog/Setup" (11) in the display "Time/Prog" (A-3)
the word "programm" appears. With the adjusting knob (1) you can chose the desired program number.
If a program number is already occupied, in the display "RPM | RCF" (A-1), the words "rotor" (M3) and "xx" (M4)
will appear. In case of free program numbers, 0 appears.
A-1
Close the lid of the centrifuge. Now proceed as described previously to set all important run parameters. If the lid
isn't closed when storing the program, in the display "RPM/RCF" (A-1), the words "FirSt" and "CLOSE Lid" (See
figure 23) flashes alternately. If you want to start the run without storing the program, in the display "RPM/RCF" (A-
1) ,the words "First" and "PrESS StoreE" (See figure 24) flashes alternately.
Figure. 23
M9
Figure. 21
M4
M3
A-3
A-3
M10
Figure. 22
11
1
6
11
1
12
Figure. 24
EN-11