The "Stop" Key; Imbalance Detection - OHAUS Frontier FC5714 Manual De Instrucciones

Ocultar thumbs Ver también para Frontier FC5714:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R

3.6.2 The "STOP" key

By the "Stop" key (10) (See figure 27) you can interrupt the run at any time. After pressing the key the centrifuge
decelerates with the respective pre-selected intensity down to stand still.

3.7 Imbalance detection

In case of the rotor not being equally loaded, the drive will turn off during acceleration. The rotor decelerates to
stand still.
When in the display "Time/Prog" (A-3) the word "error" (M11) together with the number "01" appear, the weight
difference of the samples is too large. Distribute the weight evenly.
Load the rotor as described in chapter 3.1.2 and 3.1.3.
When inside the display "Time/Prog" (A-3) the word "error" together with the number "02" (See figure 28) appear,
it could be due to the following reason: The imbalance switch is defective.
M11
A-3
Figure. 26
Figure. 27
A-3
Figure. 28
EN-13
9
1
8
10

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido