Clinician Remote Control Directions for Use

Stimulator Search and Linking

A Stimulator and Remote Control must be linked in order to communicate. A Remote Control that is
intended to be utilized by the patient must be linked as the 'Patient Remote.' The Stimulator will only
recognize one remote control device at a time as the Patient Remote; however, the Stimulator can be
linked to a second Remote Control by establishing a temporary link.
A temporary link is intended only for clinician use when the Patient Remote is not available. If a new
Remote Control device is assigned as the Patient Remote, any devices previously linked as the Patient
Remote will be cleared. Once a link between a Stimulator and Remote Control has been established,
they will remain linked unless the link is cleared.
The Clinician Menu allows you to search for a Stimulator and link to it. Stimulators that communicate
using Bluetooth Low Energy, must be placed into Pairing Mode in order to link to a Remote Control.
In order to place an IPG into Pairing Mode:
• Place the Pairing Magnet directly over the Stimulator for at least 2 seconds.
• Then remove the Pairing Magnet.
- Magnet pairing may be used on an IPG that is implanted or in the packaging.
Note: Once the magnet is removed, the Stimulator will automatically complete a device reset cycle.
This device reset cycle will take several seconds to complete. During this reset cycle, the
Stimulator will not provide stimulation and cannot communicate. Once the reset is complete,
the Stimulator will return to normal operation.
In order to place ETS into Pairing Mode:
• Press and hold
on the ETS until the battery indicator light flashes, alternating between green and
yellow.
If the Remote Control is not currently linked to a Stimulator, when the Remote Control is unlocked, the
Link to Stimulator? screen may automatically display.
From the Clinician Menu:
1. Select
Yes to link to the Stimulator. The Stimulator Search screen displays. Skip to Step 3
of this section.
Clinician Remote Control Directions for Use
92395488-02