Clinician Remote Control Directions for Use
Yes to proceed with enabling MRI Mode or
3. Select
Warning: An MRI scan is not recommended when the impedances are not within the acceptable
range. Higher or lower than normal impedances could indicate compromised Stimulator-Lead
integrity. Scanning under these conditions may increase the risk of potential adverse effects that
are listed in the "Safety Information" section of the ImageReady Full Body MRI Guidelines for
WaveWriter Alpha and WaveWriter Alpha Prime SCS Systems.
Stimulator Error Screen
If the system check fails due to a Stimulator error, MRI Mode will not be enabled and the Remote
Control will display the Stimulator Error screen. Do not perform an MRI scan if this error is displayed.
Instruct the patient to contact the healthcare provider managing their System or Boston Scientific
Technical Support.
Clinician Remote Control Directions for Use
92395488-02
40 of 1077
No to cancel the action.