Clinician Remote Control Directions for Use
ERI or EOS During MRI Mode
If MRI Mode has already been enabled and the Stimulator battery power falls below the threshold,
the Remote Control will display a message informing the patient that the Stimulator has entered the
Elective Replacement Indicator (ERI) period or has reached End of Service (EOS) of the device.
The patient can disable MRI Mode:
1. Press
to disable the MRI Mode.
2. Check the Remote Control to confirm that the Stimulator battery error message still appears.
See the "ERI (Elective Replacement Indicator) Screen (For Non-Rechargeable Stimulator Only)"
section of this manual.
Caution: The Remote Control is MR Unsafe and must not be brought into the MRI scanner
room.
Warning: Do not perform an MRI scan if MRI Mode is not enabled. Scanning under different
conditions may result in patient injury or device malfunction.
Clinician Remote Control Directions for Use
92395488-02